Category:
日本光電 Cap-ONE mainstream etCO2 semi-quantitative measurement, its
inappropriate use for continuous waveform display and the vulnerability
of the compact CO2 sensors. CardioLife TEC-7700K series, TEC-5500K series, TEC-5600 series defibrillators and ActiBiphasic output waveform.
| High inventory cost for distributors and dealers trying to sell the defibrillators |
|---|
There are four basic models of the CardioLife TEC-5600 series defibrillators and they are:
1. CardioLife TEC-5601 defibrillator
2. CardioLife TEC-5611 defibrillator
3. CardioLife TEC-5621 defibrillator
4. CardioLife TEC-5631 defibrillator
It is incomprehensible why the Pacing Unit cannot be
offered as an optional item, that it must be assembled in the factory before
delivery when other manufacturers can do it. As seen in below table, there are four rigid models to choose from. This is a lower range than the CardioLife TEC-8300K series defibrillators.
![]() |
| CardioLife TEC-5600 series comparison table |
The
rigid models implies the lack of designing skills and a burden to
users; it is impossible to upgrade a model in the field and a customer
must buy a new model if such a need arises.
![]() |
| Which of the four models should you keep stock when your competitors only need to keep one? |
| Optional connection panel with SpO2 and mainstream CO2 socket for all models |
|---|
The use of a new type yellow connector socket for mainstream CO2 is meaningless!
In the picture below,
it can be seen the connector socket for the mainstream CO2 kit set is
different from the previous model, necessitating a new set of accessory even
though the transducer is the same old one.
Note the yellow connector socket is an isolated MULTI socket because there is no Multi-parameter Unit (MPU) in the TEC-5600 series defibrillators. In other words, it is just a MULTI socket to identify the parameter code in the plug of the P903 mainstream CO2 kit set. This MULTI socket is not flexible, it only accept the mainstream CO2 kit mentioned; is this good justification for those loyal customers who want to use back their old kit set who is still in good condition?
![]() |
| A CardioLife TEC-5600 defibrillator and its side connectors |
For
those who already had bought the mainstream CO2 sensor kit for TEC-5500 series,
this is just additional expenditure with no new benefit. The
change to using a yellow MULTI socket has negative captured value for such loyal customers.
![]() |
| The P903 on the left is for TEC-5600 series while the P906 on the right is for TEC-5500 series. The two are identical except for the connector plug. |
The
CardioLife TEC-5600 series defibrillators have no reason to use a new type socket different from that of the TEC-5500 series.
| Why you should watch out the careless and dangerous use of estimated CO2 values as true CO2 values |
|---|
The cap-ONE TG-920P CO2 sensor kit (P907) has very small sensors because of the adoption of semi-quantitative measurements. The method is
not commonly seen and many are not aware of the risks of CO2
readings from the semi-quantitative CO2 kit sets; to make matter
worse, the semi-quantitative measurements are also being fed to
display a flawed continuous CO2 waveform on screen to users.
![]() |
| Nihon Kohden cap-ONE P907 (TG-920P) mainstream CO2 sensor kit |
Shown below is another TG-900P etCO2 kit set (order code P903) that makes semi-quantitative CO2 measurements on a traditional mainstream CO2 sensor. The TG-901T3 kit set (order code P906)
is the same thing but using a non-coded connection plug. The medical
devices from same manufacturer that make use of semi-quantitative CO2
kit sets for patient CO2 measurements and waveform include:
- Life Scope patient monitors
- Vismo patient monitors
- Cap-STAT OLG-2800
- CardioLife defibrillators
- Neurofax EEG machines etc.
![]() |
| Nihon Kohden semi-quantitative CO2 kit sets with traditional mainstream transducer |
| Semi-quantitative CO2 measurements are not cheap alternatives to quantitative measurements |
|---|





